Its recommended that you use C0G type ceramic capacitors in the signal path to reduce signal distortion
- For the ECG path this includes CHPF, CDME, CCMEP, CCMEN
- For the BioZ path this includes CDRVP, CDRVN, CDMB, CCMBP, CCMBN