Article What type of capacitors are recommended for use with the MAX30001 to minimize the noise?

Its recommended that you use C0G type ceramic capacitors in the signal path to reduce signal distortion

  • For the ECG path this includes CHPF, CDME, CCMEP, CCMEN
  • For the BioZ path this includes CDRVP, CDRVN, CDMB, CCMBP, CCMBN