This is critical. Unfortunately, the user will need this 1nF cap to keep the PLL stable. It cannot be grounded. Please try to rework the board by adding 1nF bypassed to Ground.
It is optional. It is required if the purpose of the device is for medical purposes.
If the ECG feature is only used for non-medical purpose such as Bio-Identification, then FDA approval is not needed…
BioZ Reading (ohm) = (VREF×[20 bit ADC code]) / (BioZ Drive current × 2^19 × BioZ Gain Setting)
Where VREF = 1V. BioZ Drive Current is programmable to 8uA, 16uA, 32uA, 48uA, 64uA, 80uA, or 96uA. BioZ…
Question:
As per MAX30001 datasheet, the AFE could also be used to measure hydration. Customers want to understand how this could be achieved using MAX30001? Do they need external analog sensor?
…
Since ECG electrodes would not be run in parallel over their entire path from electrode to the module, one approach to ensure both ECG inputs “see” the same noise is to add an antenna that runs in parallel…
Maxim ECG AFE (MAX30001) is a complete biopotential and bio-impedance analog front end solution for wearable consumer products. In addition to the AFE amplifier and ADC commonly found in competitive parts…
The 100pF shunting capacitors will not cause a significant problem. They will tend to reduce the signal received back from the body a bit, but not substantially.
It certainly shouldn’t be a problem for…
The demodulator signal leads the drive current signal in phase. The phase by which the demodulator leads the drive current signal is set by the bits BIOZ_PHOFF[3:0] in CNFG_BIOZ (0x18).
The electrodes can be placed on the back (ensuring that you are measuring the same vector). However placing the electrodes on the back will decrease the signal amplitude and may result in unwanted movement…