1. When using the MAXREFDES100 to measure ECG, the resistors R6, R21, R44, and R45 need to be removed first.
2. If that's done and the issue persists, please make sure that the ECGP and ECGN switches are connected in the GUI.
1. When using the MAXREFDES100 to measure ECG, the resistors R6, R21, R44, and R45 need to be removed first.
2. If that's done and the issue persists, please make sure that the ECGP and ECGN switches are connected in the GUI.