No, the GUI of the MAX86150EVSYS and a PC must transmit data and communicate via Bluetooth. The USB interface is only a power supply and charging interface, not a data communication interface.
Yes, if a Micro USB is connected to the MAX32630FTHR while a battery is not connected, the MAX86150EVSYS will operate correctly. The Micro USB is used to charge the battery if it's depleted once connected…
The MAX86150 was designed for on-demand applications, not continuous applications (typically medical). This IC was intended for use in cellphones. The following device performs reflective simulation of…
The ECG electrodes, TP1 and TP2, on the MAX86150EVSYS# PCB are made from Stainless Steel Sheets which are custom fabricated to fit on the MAX86150EVSYS#.
In order to turn off the MAX86150EVSYS, you must hold down SW1 on the MAX32630FTHR for 12 seconds. This shutsdown the MAX14690 PMIC being used for power regulation in the system.
The ECG filter library source code (GCC compiler) is available after filling out a SLA form.
Kindly request the release of the source code by creating a case on the technical support portal
https…