After the configuration register is set, Mode configuration should remain unchanged unless it is reset. So if the data varies, the device must have been reset.
The LEDs in the MAX30102 are not optically altered in any way to vary the transmission angle. We expect the LEDs to transmit homogeneously out of their encapsulation.
While the MAX30102 is an optical sensor which can be used for Blood Glucose monitoring applications, it cannot be used for non invasive blood glucose monitoring as optical-reflectometry method requires…
The temperature sensor on board the MAX30102 is not needed in Maxim's HRM/SpO2 algorithms. The temperature sensor was leftover from an older through-process. It is not an accurate temperature sensor so…
We cannot guarantee 3.0V will work for VLED+ in all parts if you use both IR and Red LEDs. There are part to part variations on the nominal LED forward voltages that may cause things not to work at 3.0V…
Answer: No, the photodiode is optimized for the spectrum of the LEDs used. (The IR LED is 880 nm; the red LED is 660 nm.) Measuring body temperature requires a good response to wavelengths more than 10…
As per simulators, this is one of the devices we are aware of that does Reflective simulation of PPG signals ( https://www.whaleteq.com/en/product.php?act=view&id=20 ).
It is recommended to add an external cover glass. The cover glass must guarantee over 90% transmission of red and infrared light. Otherwise, the signal quality will degrade and affect the performance of…
Yes, it does. Please find below the test procedure and the scopeshot that reflects this. Hardware used : MAX32600MBED and MAXREFDES117, which has one MAX30102 on the board. Test procedure MAX30102 is set…