ADIS16488A Datasheet Error: FIR Filter Configuration

Document created by NevadaMark Employee on Jan 2, 2018
Version 1Show Document
  • View in full screen mode
DATASHEETS

ADIS16488A (Rev E, page 23, left column, second paragraph)

URL

http://www.analog.com/media/en/technical-documentation/data-sheets/ADIS16488A.pdf#Page=23

PROBLEM DESCRIPTION

 

One of the codes in the example configuration sequence is not correct. Please see the following graphic for the precise location of the incorrect code.

SOLUTION

When using this reference, replace the 0x9657 code (see following graphic for specific location) with 0x962F

RESOLUTION PLAN

Revision F of this datasheet will address this problem.

RESOLUTION STATUS

OPEN

Attachments

    Outcomes