IMU Evaluation Software Bug, ADIS16265, Data Capture 

Document created by NevadaMark Employee on Feb 3, 2017Last modified by NevadaMark Employee on Feb 3, 2017
Version 3Show Document
  • View in full screen mode
TITLEIMU Evaluation Software Bug, ADIS16265, Data Capture
SOFTWARE PACKAGEIMU Evaluation Software
VERSIONv1.17.5
SYSTEMSEVAL-ADIS, EVAL-ADIS2
DEVICE UNDER TESTADIS16265/PCBZ
OTHER PRODUCTS THAT COULD BE AFFECTED
ADIS16260/PCBZ (obsolete)
DESCRIPTION

The sample update rate in the ADIS16265 is 2048SPS, when SMPL_PRD[7:0] = 0x00.  The Data Capture window is reporting this 512.033 SPS, which is incorrect. It appears like this error is not influencing the data acquisition rate, as a 8192 sample record took approximately 4 seconds to complete. 

SOLUTIONFor the short term, please ignore this incorrect sample rate label.  We anticipate fixing this in an upcoming revision, sometime in the next 2-3 weeks. 
1 person found this helpful

Attachments

    Outcomes