4Wire-Isolated-Bioimpedance ADuCM350 Connections To Sensor

Document created by LiamR Employee on Apr 24, 2014Last modified by LiamR Employee on Apr 30, 2014
Version 3Show Document
  • View in full screen mode

4-Wire-Bio-Config Board

 

  • By default used to measure a matrix of ideal R and C components
  • To measure “real world” sensors modifications need to be made to board.
  • Please refer to 4-Wire Bio Config 2 Layout and Schematic in Design Documentation folder when evaluating and modifying board.

 

 

Step 1: Validate Board with Software

  • Measure 20K (0.1% accurate) resistor
    • Insert following Links:LK7,LK8,Lk9,LK10,LK12 and LK19.
  • Measure 20K + 220pF(5% accurate) capacitor
    • Insert following Links:LK7,LK8,Lk9,LK10,LK12 and LK21.

Step 2: Electrode Hook Up

  • Using the “4-Wire Bio Config” board remove the RC components that replicate the sensor and replace with real sensor.
    • Rlimit resistor and Isolation capacitors will need to be maintained in signal chain to meet IEC standards.
    • Raccess will need to be removed from measurement.
  • Insert following Links:LK7,LK8,Lk9,LK10.

 

4-Wire-Isolated Electrode Configuration

 

Block Diagram

Insert following Links: LK7, LK8, LK9 and LK10. All other links on board are removed.

 

Picture6.jpg

 

Schematic

Because Sensor Links are opened no current flows through Raccess on board.

 

 

Picture7.jpg

Layout

 

Picture8.jpg

Board Pictures

 

Picture9.jpg

Picture10.jpg