Search FAQs on the left to see if your question has been answered. Click on the dropdown to view all of the documents associated with the product. If you can't find your question, click on Ask a Question

Labview Error 1172 (.NET exception) when starting AD5780 evaluation software

I get an error 1172 when starting the evaluation software. I am using an SDP-S
control board with the AD5780 evaluation kit.

 

Error 1172 is a generic LabVIEW error indicating there was an error thrown
during a .NET assembly call. The sdpAPI1.dll is a .NET assembly. I expect this
is because you are using the incorrect SDP board. 
The SDP-B (EVAL-SDP-CB1Z) and not the SDP-S is the required controller board
for the AD5780 evaluation system.

The connector on the SDP-B and SDP-S is the same so physically the boards
(controller and Eval) might connect, however the SDP-S doesn’t have the same
number of peripherals available as the SDP-B, so a lot of the pins are No
Connects.