Article AD5522: Voltmeter Mode

Can AD5522 be used in voltmeter mode and how is it done?

Yes, it can be used in voltmeter mode. Set FORCE1 bit to 1 to put the force amp in High-Z mode.

  1. The Force function for Channel 3 should be placed in High-Z FOH voltage mode(FORCE[1:0]=0x2) while its CH_EN should be 1 and change the MEASOUT GAIN=1. Before pressing the update button, please make sure to remove LK10,11,12 and 13 so as not to damage the on board ADC.

 

  1. 1 Volt should be seen at the Measout of channel 3.

 

Can one AD5522 channel be used as source and another channel as voltmeter?

Yes, the diagram below shows channel 1 as source and another channel 2 as a voltmeter.

 

Please take note that FOH1 pin should always be shorted to MEASVH1 to close the active loop for this channel and give out a constant voltage.

  1. The Force function for Channel 3 should be placed in High-Z FOH voltage mode (FORCE[1:0]=0x2) while its CH_EN should be 1. Please check that WRITE=0x831B300 before pressing WRITE PMU REGISTER.

 

 

  1. The force function for Channel 1 should be in Force Voltage mode(FORCE[1:0]=0x0) while its CH_EN should be 1. Please check that WRITE=0x221B300 before pressing WRITE PMU REGISTER.

 

  1. Force a 1 volt(0x8B60) out of the channel 1 FOH and change the MEASOUT GAIN=1. Before pressing the update button, please make sure to remove LK10,11,12 and 13 so as not to damage the on board ADC.

 

  1. 1 Volt should be seen at the Measout of channel 3.