How to reduce the transient response when switching between external and internal current range?
When switching between the internal and external current ranges, one must bear in mind that force amplifiers within AD5522 either get activated or inactivated. In that process, it is likely for transient responses to occur at the FORCE output.
The below oscilloscope shot shows the transient response of the FOH voltage with 1kOhm load when switched from the 2mA current range to the external current range.

However, there are ways to reduce the amplitude of transient response:
Enabling the External Range Always On mode
- This is the recommended method of reducing the force output glitches as this doesn’t require disabling the force amplifier or putting its output on high-Z. Also, since this method already has the external amplifier always active, this also results in much greater glitch optimization vs. the first two methods.
- Be aware of the additional power consumption.
- This is also the most effective way of reducing the magnitude of transient responses.
It is noticeable from the below plot that the voltage transient has been almost flattened:

In any case the additional power consumption cannot be afforded, there are some alternative ways to optimize AD5522’s glitch response. It should be noted, however, that the amounts of transient response optimization using the following methods do not compare with the first recommended solution.
Disabling the channel prior to current range change.
- This method requires enabling the channel again after the current range change.
Putting the force amplifier to high-Z before changing current ranges
- This method requires reverting back to the original force setting of the device after the range change.